Evaluation of surface image guidance and Deep inspiration Breath Hold technique for breast treatments with Halcyon
Related Experiment Video
Updated: Aug 5, 2025

Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Frederik Crop1, Julien Laffarguette1, Ilias Achag1
1Medical Physics, Centre Oscar Lambret, Lille, France.
Purpose:
To evaluate the accuracy/agreement of a three-camera Catalyst Surface Guided Radiation Therapy (SGRT) system on a closed-gantry Halcyon for Free-Breathing (FB) and Deep Inspiration Breath Hold (DIBH) breast-only treatments.
Methods:
The SGRT positioning agreement with Halcyon couch and cone-beam computed tomography (CBCT) was evaluated on phantom and by evaluation of 2401 FB and 855 DIBH breast-only treatment sessions. The DIBH agreement was evaluated using a programmable moving support. Dose agreement was evaluated for manual SGRT-assisted beam interruption and Halcyon arc beam interruption.
Results:
Geometrical phantom agreement was < 0.4 mm. Couch and SGRT agreement for an anthropomorphic phantom resulted in 95% limits of agreement in Right-Left/Feet-Head/Posterior-Anterior (RL/FH/PA) directions of respectively ± 0.4/0.8/0.5 mm and ± 1.1/1.1/0.6 mm in the virtual and real isocenter. FB-SGRT-assisted patient positioning compared to CBCT positioning resulted in RL/FH/PA systematic differences of -0.1/0.1/2.0 mm with standard deviations of 2.7/2.8/2.4 mm. This mean systematic difference had three origins: a) couch sag/isocenter difference of ≤ 0.5 mm. b) Average reconstructed FB-CBCT images do not visually represent the average respiratory position. c) CBCT-based positioning focused on the inner thoracic interface, which can introduce a mean positioning difference between SGRT and CBCT. Manual SGRT-assisted beam interruption and arc interruptions resulted in mean gamma passing rates > 97% (0.5%/0.5 mm) and mean absolute differences < 0.3%.
Conclusions:
Accuracy was comparable with breast-only C-arm SGRT techniques, with different tradeoffs. Depending on the patient's morphology, real-time tracking accuracy in the real isocenter can be reduced. This study demonstrates possible discordances between SGRT and CBCT positioning for breast.
Related Concept Videos
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...