Integrating Augmented Reality Tools in Breast Cancer Related Lymphedema Prognostication and Diagnosis
Published on: February 6, 2020
6.7K
Three-Dimensional Surface Analysis for Preoperative Prediction of Breast Volume: A Validation Study
Carrie S Stern1, Ethan L Plotsker1, Robyn Rubenstein1
1From the Plastic and Reconstructive Surgery Service, Department of Surgery, Memorial Sloan Kettering Cancer Center.
Plastic and Reconstructive Surgery
|March 30, 2023
Summary
Three-dimensional imaging with VECTRA Analysis Module (VAM) offers more accurate breast volume prediction for mastectomy planning than VECTRA Body Sculptor (VBS). Patient characteristics influence prediction accuracy, requiring careful consideration by surgeons.
Area of Science:
- Medical Imaging
- Plastic Surgery
- Oncology
Background:
- Preoperative breast volume prediction is crucial for breast reconstruction, patient education, and risk stratification.
- Few studies have validated the accuracy of preoperative 3D surface imaging for predicting breast volume.
Purpose of the Study:
- To evaluate the accuracy of VECTRA XT three-dimensional imaging, specifically VECTRA Analysis Module (VAM) and VECTRA Body Sculptor (VBS), in predicting mastectomy breast volume.
- To compare the predictive accuracy of VAM and VBS against intraoperative mastectomy specimen weights.
Main Methods:
- A retrospective review of 179 patients (266 breasts) who underwent mastectomy with preoperative VECTRA XT 3D imaging from 2020-2021.
- Volumetric analysis performed using VECTRA Analysis Module (VAM) and VECTRA Body Sculptor (VBS) based on standard anatomical breast borders.
- Predictive accuracy defined as estimates within ±10% or ±100 g of the intraoperative mastectomy specimen weight.
Main Results:
- No significant difference was found between mean mastectomy weight (620.8 ± 360.3 g) and mean VAM estimate (609.5 ± 361.9 g).
- Mean VBS estimate (498.9 ± 337.6 g) significantly differed from mean mastectomy weight (P < 0.001).
- Predictive accuracy within ±100 g was achieved by 58.7% of VAM estimates and 44.4% of VBS estimates. Body mass index, body surface area, and ptosis grade impacted prediction accuracy.
Conclusions:
- VECTRA Analysis Module (VAM) demonstrates higher accuracy in predicting mastectomy weight compared to VECTRA Body Sculptor (VBS).
- Discrepancies may arise from differences between surgical mastectomy borders and the breast borders used in volumetric analysis.
- Surgeons should integrate patient physical characteristics with 3D imaging data for improved preoperative breast volume prediction.