Long-term ECMO, efficiency and performance of EUROSETS adult A.L.ONE ECMO oxygenator
Ignazio Condello1, Roberto Lorusso2,3, Giuseppe Nasso4
1Department of Cardiac Surgery, Anthea Hospital, GVM Care & Research, Via Camillo Rosalba 35/37, 70124, Bari, Italy. ignicondello@hotmail.it.
Background:
The management of the oxygenator can be prolonged in the long-term procedures especially during extracorporeal membrane oxygenation (ECMO) for bridge to transplant or bridge to recovery. Long-term use often involves an overrun of the time of use with respect to certification of the oxygenating module of 14 days, for the maintenance of performance and efficiency of the oxygenator. The evaluation of the long-term oxygenator efficiency is complex and depends on the: patient pathology, ECMO configuration, the management of coagulation and anticoagulation, materials selection and circuit components, the structure, design and performance of the oxygenator. In this context we investgated the long-term performance of the A.L.ONE Eurosets ECMO oxygenator in relation to the parameters prodromal to replacement.
Methods:
We retrospectively collected eight years data from Anthea Hospital GVM Care & Research, Bari, Italy on the long-term use exceeding 14 days of Eurosets A.L.ONE ECMO Adult oxygenator in Polymetylpentene fiber, for ECMO procedures, including the procedures: Veno Arterial (VA) ECMO post-cardiotomy or not, veno-venous (VV) ECMO. The primary end points were the evaluation of Gas Transfer: oxygen partial pressure (PO2) post oxygenator, Carbon dioxide partial pressure (PCO2) post oxygenator, the oxygen transfer across the oxygenator membrane V'O2, differential CO2 content across oxygenator; Pressure monitoring: oxygenator pressure Drop in relation to Blood flow rate (BFR) (ΔP); Hematologic values: Hemoglobin, Fibrinogen, Platelets, aPTT, D-Dimer, LDH.
Results:
Nine VA ECMO patients who used the oxygenator for 18.5 days and two VV ECMO patients who used the oxygenators for 17.2 days on the seventeenth days reported average values PaO2 (267 ± 29 mmHg); PaCO2 (34 ± 4 mmHg) with gas blender values set to 3.8 ± 0.6 L/min of air and a FiO2 of 78 ± 5%; the transfer across the oxygenator membrane V'O2 was 189 ± 43 (ml/min/m2). The mean peak of partial pressure of carbon dioxide from the gas exhaust of oxygenator (PECO2) was 38 ± 4 mmHg; differential CO2 across the oxygenator "pre-oxygenator PCO2-post-oxygenator PCO2" (18 ± 6 mmHg); the mean blood flow rate (BFR) 4.5 ± 0.6 (L/minute); the pump revolution per minutes mean maximum rate was 4254 ± 345 (RPM); the mean pressure drop (ΔP) was 76 ± 12 mmHg; the mean peak of d-dimers (DDs) was 23.6 ± 0.8 mg / dL; the mean peak of LDH was 230 ± 55 (mg/dl); fibrinogen mean peak 223 ± 40 (mg/dl).
Conclusions:
The performance of the Eurosets A.L.ONE ECMO Adult polymethylpentene fiber oxygenator in our experience has proven efficiency in terms of O2 uptake and CO2 removal, blood fluid dynamics, metabolic compensation and heat exchange in the long-term treatment. The device was safe without iatrogenic problems over a period of 14 days in the patients undergoing ECMO VA and in all patients undergoing VV ECMO with continuous administration of anticoagulation therapy.
More Related Videos
09:54Lung Rapid Recovery Procurement Combined with Abdominal Normothermic Regional Perfusion in Controlled Donation after Circulatory Death
Published on: August 15, 2022
09:31Normothermic Negative Pressure Ventilation Ex Situ Lung Perfusion: Evaluation of Lung Function and Metabolism
Published on: February 14, 2022
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation I: Indication and Settings
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: