Establishing the Ideal Conditions to Create an Airway Fire Using a Porcine Airway Model
Andrew M Bysice1, Tyler Oswald2, Luis E Mendoza Vasquez2
1Department of Otolaryngology-Head and Neck Surgery, Schulich School of Medicine & Dentistry, London Health Sciences Centre Western University London Ontario Canada.
Objective:
Airway fires are a rare but devastating complication of airway surgery. Although protocols for managing airway fires have been discussed, the ideal conditions for igniting airway fires remain unclear. This study examined the oxygen level required to ignite a fire during a tracheostomy.
Study Design:
Porcine Model.
Setting:
Laboratory.
Methods:
Porcine tracheas were intubated with a 7.5 air-filled polyvinyl endotracheal tube. A tracheostomy was performed. Monopolar and bipolar cautery were used in independent experiments to assess the ignition capacity. Seven trials were performed for each fraction of inspired oxygen (FiO2): 1.0, 0.9, 0.7, 0.6, 0.5, 0.4, and 0.3. The primary outcome was ignition of a fire. The time was started once the cautery function was turned on. The time was stopped when a flame was produced. Thirty seconds was used as the cut-off for "no fire."
Results:
The average time to ignition for monopolar cautery at FiO2 of 1.0, 0.9, 0.8, 0.7, and 0.6 was found to be 9.9, 6.6, 6.9, 9.6, and 8.4 s, respectively. FiO2 ≤ 0.5 did not produce a flame. No flame was created using the bipolar device. Dry tissue eschar shortened the time to ignition, whereas moisture in the tissue prolonged the time to ignition. However, these differences were not quantified.
Conclusion:
Dry tissue eschar, monopolar cautery, and FiO2 ≥ 0.6 are more likely to result in airway fires.