In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
Mobile bearing orbit on the tibial component in Oxford unicompartmental knee arthroplasty
Yoshihito Suda1, Takafumi Hiranaka2, Tomoyuki Kamenaga1
1Department of Orthopaedic Surgery and Joint Surgery Centre, Takatsuki General Hospital, Osaka, Japan; Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, Hyogo, Japan.
Purpose:
To evaluate the bearing orbit of the tibial component during extension-flexion motion in Oxford unicompartmental knee arthroplasty.
Materials And Methods:
A total of 32 knees in 25 patients with medial osteoarthritis who underwent Oxford unicompartmental knee arthroplasty were evaluated. The distance between the vertical wall of the tibial component and the bearing (wall-bearing distance) and that between the anterior edge of the tibial component and the bearing (sagittal bearing position) were measured at 0°, 30°, 60°, 90° and 120° knee flexion with neutral tibial rotation (extension-flexion motion), and internal and external tibial rotation with 90° knee flexion (tibial rotation motion). A custom-made rounded trial bearing and caliper were used for this measurement. We calculated the wall-bearing distance, change in extension-flexion motion and tibial rotation motion. Wall-bearing distances and change in wall-bearing distance were compared using ANOVA or t-test.
Results:
The wall-bearing distance was smallest at 60° and increased 1.0 ± 1.1 mm in knee extension and 1.1 ± 1.5 mm in knee flexion. The bearing moved posteriorly with knee flexion, and the sagittal bearing position increased by 8.1 ± 3.4 mm during extension-flexion motion. Consequently, the bearing moved in a rough C-shaped orbit of the tibial component.
Conclusions:
The mobile bearing moves in a rough C-shape and is mostly close to the vertical wall of the tibial component at 60°. The wall-bearing distance can change during extension-flexion motion and might be influenced by tibial component rotation. To avoid bearing separation from or contact with the vertical wall that may cause bearing dislocation, the wall-bearing distance should be evaluated before keel slot preparation.
Related Concept Videos
Pivot Bearings
A pivot bearing is a specialized type of bearing designed to support axial loads on a rotating shaft. The bearing surface, or the pivot, is positioned at the end of a shaft to support the axial thrust. The pivot may...
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Ankle Joint
Bones of the Lower Limb: Tibia and Fibula
Bones of the Lower Limb: Femur and Patella
Eccentric Axial Loading in a Plane of Symmetry