Related Experiment Video
Updated: Aug 4, 2025

Fully Endoscopic Mitral Valve Repair with Percutaneous Cannulation of Groin Vessels
Published on: May 26, 2023
Safety and Effectiveness of a Sequential Suture and Plug Vascular Closure Devices Technique for Large-Bore Access
Chen Xu1, Jiang Li1, Jian-Ting Zhao1
1Department of Vascular Surgery and Intervention, the Affiliated Suzhou Hospital of Nanjing Medical University, Gusu District, Suzhou, People's Republic of China.
Purpose:
To evaluate the safety and effectiveness of sequential sutures and plugged vascular closure devices (VCDs) for large-bore access closure during percutaneous access endovascular aneurysm repair (PEVAR).
Materials And Methods:
Data on 16 patients who underwent PEVAR at the authors' center from January 2022 to May 2022 were retrospectively reviewed. The median age was 72 years (interquartile range [IQR], 59-75 years), with a male-to-female ratio of 3:1. All patients received sequential suture and plug VCDs using dual Exoseal after 1 Proglide for access closure. Success was defined as the ability to achieve complete hemostasis and was confirmed by ultrasonography. The patients were followed up for access-related adverse events at 30 and 90 days after the procedure, and the severity was graded according to the Society of Interventional Radiology (SIR) classification.
Results:
Overall, 24 access sites were included. The median sheath size was 21 F (IQR, 18-23 F). The median hemostasis time was 11.0 minutes (IQR, 9.3-13.0 minutes), the median procedural time was 133.5 minutes (IQR, 102.5-151.0 minutes), and the median length of stay was 5 days (IQR, 4.0-6.8 days). The success rate was 95.8%, and a pseudoaneurysm (SIR Grade 2) developed in 1 patient, which was treated by a percutaneous injection of thrombin. No other access-related adverse events occurred, and the total adverse event rate was 4.2%.
Conclusions:
Placement of sequential suture and plug VCDs using 1 Proglide and dual Exoseal is a safe and effective method and may be an option for access closure during PEVAR.
More Related Videos
07:21Microsurgical Creation of Giant Bifurcation Aneurysms in Rabbits for the Evaluation of Endovascular Devices
Published on: September 8, 2023
08:05Occlusion of the Great and Small Saphenous Vein Using Copolymeric Glue Based on N-Butyl Cyanoacrylate and Methacryloxy Sulfolane
Published on: December 9, 2022
Related Concept Videos
Aneurysm III: Interprofessional Care
Aneurysm IV: Nursing Management