Validation and application of volumetric absorptive microsampling (VAMS) dried blood method for phenylalanine
Lan Gao1, Neil Smith1, Diksha Kaushik1
1PTC Therapeutics, Inc., South Plainfield, NJ 07080, USA.
Background:
Frequent blood phenylalanine (Phe) measurement is required for phenylketonuria (PKU) patients for diagnosis and disease status monitoring. Though various methods are available for blood Phe measurement, there is a lack of validated quantitative methods for measuring Phe with less than 15% variability. A method to allow at home blood sample collection for the PKU community is in high demand.
Methods:
A volumetric absorptive microsampling (VAMS) dried blood collection high-performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS) method was developed and fully validated for blood Phe measurement in compliance with regulatory guidances. The method accuracy, precision, stability, selectivity, matrix and hematocrit effects were assessed. A venous plasma collection HPLC-MS/MS method was developed and validated as a reference method. 311 matching VAMS and plasma samples were collected from 24 PKU subjects in a Phase 2 clinical study. Phe measurements using the two methods were compared.
Results:
Both VAMS and the plasma sample collection methods met the acceptance criteria for Good Laboratory Practice (GLP) bioanalytical analysis. Comparisons showed a high Pearson's correlation of 0.9813. The Passing-Bablok analysis showed that the difference was estimated to be less than 5% and Bland Altman analysis indicated that the difference was proportional with Phe concentration and for the majority of samples (88.85%) the measurement was within ±20% difference. Following 7 days treatment with 60 or 20 mg/kg/day PTC923 (Sepiapterin) or 20 mg/kg/day sapropterin, PKU patients exhibited respectively -206.4, -146.9, and -91.5 µmol/L reductions of blood Phe as measured by the VAMS method.
Conclusions:
Concordant results were obtained using VAMS and plasma methods, which demonstrated that VAMS is a reliable method for clinical applications to monitor blood Phe for PKU patients.