Video Head Impulse Test: A Comparison and Analysis of Three Recording Systems
Related Experiment Video
Updated: Aug 4, 2025

Testing of all Six Semicircular Canals with Video Head Impulse Test Systems
Published on: April 18, 2019
Mercy George1, Arif Ali Kolethekkat2, Perez Yoan1
1ENT Service, Centre Hospitalier Universitaire Vaudois (CHUV), Rue du Bugnon, 46, 1011 Lausanne, Switzerland.
Abstract:
The present study aimed to compare the efficacy and reproducibility of three existing systems in a group of healthy individuals for vHIT. A prospective randomized study was conducted on 12 healthy individuals. The vHIT tests were carried out. The values of the gains for the 3SCCs of each ear were collected using the three devices. The expected average gain of 1 was the gain standard. The statistical significance in the difference in gains were assessed. There is good reproducibility of the results of the vHIT examination. EyeSeeCam appeared to be the worst performing system with a slightly overstated average gain of 1.15. Otometrics has longest average examination time per patient. Synapsis is the system with the best quality/time invested and easiest to access. Video head impulse system depends on the examiner's preference, reproducible and superimposable depending on the examiner and experience.

