Correlation Between Lens Density Measured by Swept-Source Optical Coherence Tomography and Phacodynamic Parameters of
Xueer Wu1,2,3, Lulu Chen1,2,4, Zhangliang Li1,2,3
1Eye Hospital and School of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, Zhejiang, China.
Purpose:
To evaluate the correlation between lens density measured by IOL-Master 700 based on swept-source optical coherence tomography (SS-OCT) technology and the phacodynamic parameters of Centurion phacoemulsification in cataract surgery.
Methods:
This prospective study included 66 patients (83 eyes) with age-related cataracts. Using the Lens Opacities Classification System III (LOCS III), the lens nuclear color (NC), lens nuclear opalescence (NO), cortical (C), and posterior subcapsular (P) opacities were obtained. Six meridian orientations of IOL-Master 700 images were captured, and the lens and nuclear regions were analyzed using ImageJ to generate the average lens nucleus density (AND) and average lens density (ALD). Phacodynamic parameters were recorded. The correlation between lens density and the phacodynamic parameters was analyzed. According to the AND, patients were divided into four groups (soft, medium-hard, hard, and extremely hard nucleus), and the phacodynamic parameters were compared among groups.
Results:
The correlation between the AND obtained by LOCS III grading and SS-OCT-based cataract quantification system score (NC and NO) was statistically significant (rNC = 0.795, rNO=0.794, both p = .000). AND correlated significantly with cumulative dissipated energy (CDE, r = 0.545, p = .000), total ultrasound time (TUST, r = 0.354, p = .001), and total torsional ultrasound time (TTUT, r = 0.314, p = .004). Among the four groups divided by AND, the difference in CDE (P13 = 0.002, P14 < 0.001, P24 = 0.002) was statistically significant.
Conclusion:
AND measured by IOL-Master 700, SS-OCT correlated significantly with LOCS III classification and phacodynamic parameters of the Centurion system, especially with CDE, TUST, and TTUT. AND can be used as an indicator for quantitative evaluation and help inform the surgical plan.