Accuracy and Precision of Iodine Quantification in Subtracted Micro-Computed Tomography: Effect of Reconstruction and
Lízbeth Ayala-Dominguez1,2, Luis-Alberto Medina3,4, Carmen Aceves5
1Departamento de Física Experimental, Instituto de Física, Universidad Nacional Autónoma de México, Circuito de La Investigación Científica, Ciudad Universitaria UNAM, Mexico City, 04510, Mexico. lizbeth.ayaladominguez@wisc.edu.
Purpose:
To evaluate the effect of reconstruction and noise removal algorithms on the accuracy and precision of iodine concentration (CI) quantified with subtracted micro-computed tomography (micro-CT).
Procedures:
Two reconstruction algorithms were evaluated: a filtered backprojection (FBP) algorithm and a simultaneous iterative reconstruction technique (SIRT) algorithm. A 3D bilateral filter (BF) was used for noise removal. A phantom study evaluated and compared the image quality, and the accuracy and precision of CI in four scenarios: filtered FBP, filtered SIRT, non-filtered FBP, and non-filtered SIRT. In vivo experiments were performed in an animal model of chemically-induced mammary cancer.
Results:
Linear relationships between the measured and nominal CI values were found for all the scenarios in the phantom study (R2 > 0.95). SIRT significantly improved the accuracy and precision of CI compared to FBP, as given by their lower bias (adj. p-value = 0.0308) and repeatability coefficient (adj. p-value < 0.0001). Noise removal enabled a significant decrease in bias in filtered SIRT images only; non-significant differences were found for the repeatability coefficient. The phantom and in vivo studies showed that CI is a reproducible imaging parameter for all the scenarios (Pearson r > 0.99, p-value < 0.001). The contrast-to-noise ratio showed non-significant differences among the evaluated scenarios in the phantom study, while a significant improvement was found in the in vivo study when SIRT and BF algorithms were used.
Conclusions:
SIRT and BF algorithms improved the accuracy and precision of CI compared to FBP and non-filtered images, which encourages their use in subtracted micro-CT imaging.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Redox Titration: Iodimetry and Iodometry
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...