Detection of respiratory activity in newborn infants using a noncontact vision-based monitor
Caillin Eastwood-Sutherland1, Kathleen Lim2, Timothy J Gale1
1School of Engineering, University of Tasmania, Hobart, Tasmania, Australia.
Objective:
To examine the effectiveness of a noncontact vision-based infrared respiratory monitor (IRM) in the detection of authentic respiratory motion in newborn infants.
Study Design:
Observational study in a neonatal intensive care unit.
Methods:
Eligible infants lay supine with torso exposed under the IRM's infrared depth-map camera and torso images were recorded at 30 frames/s. Respiratory motion waveforms were subsequently derived from upper (IRMupper ) and lower (IRMlower ) torso region images and compared with contemporaneous impedance pneumography (IP) and capsule pneumography (CP). Waveforms, in 15 s investigative epochs, were scanned with an 8 s sliding window for authentic respiratory waveform (spectral purity index [SPI] ≥ 0.75, minimum five complete breaths). Maximum SPI and frequency of occurrence of authentic respiratory waveform in 15 s epochs were compared between monitoring modalities in pooled and per patient data (Friedman ANOVA).
Results:
Recordings comprised 532 min of images from 35 infants, yielding 2131 investigative epochs, with authentic respiratory motion detected in all infants. For CP, IP, IRMupper , and IRMlower , the proportion of epochs containing authentic respiratory motion in pooled data were 65%, 50%, 36%, and 48%, with median SPImax of 0.79, 0.75, 0.70, and 0.74, respectively. Per-patient average SPImax was 0.79, 0.75, 0.69, and 0.74 for CP, IP, IRMupper , and IRMlower with proportion of authentic respiratory motion being 64%, 50%, 29%, and 49%, respectively.
Conclusion:
An IRM focused on the lower torso detected authentic respiratory motion with comparable performance to IP in newborn infants in intensive care and deserves further investigation.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...