Agreement Between Viscoelastic Coagulation Monitor (VCM), TEG 5000, and Coagulation Tests in Critically Ill Patients:
Mauro Panigada1, Andrea Meli1, Andrea Forastieri Molinari2
1From the Department of Anesthesia and Critical Care, Fondazione IRCCS Ca' Granda Ospedale Maggiore Policlinico, Milan, Italy.
Summary
The viscoelastic coagulation monitor (VCM) and TEG 5000 (TEG) show different responses to heparin in critically ill patients. VCM-MCF effectively reflects platelet count, even with unfractionated heparin.
Area of Science:
- Critical Care Medicine
- Hematology
- Biomedical Engineering
Background:
- Viscoelastic coagulation monitoring is crucial for managing critically ill patients.
- Comparing the performance of Viscoelastic Coagulation Monitor (VCM) and TEG 5000 (TEG) is essential for clinical application.
- The influence of heparin on these devices requires further investigation.
Purpose of the Study:
- To evaluate the agreement between VCM and TEG parameters.
- To assess the relationship between VCM/TEG parameters and standard coagulation tests.
- To determine the impact of heparin (unfractionated and low molecular weight) on VCM and TEG.
Main Methods:
- Multicenter study involving 127 critically ill patients.
- Simultaneous analysis of Viscoelastic Coagulation Monitor (VCM), TEG 5000 (TEG), and laboratory samples.
- Agreement assessed using Bland and Altman plots; correlations analyzed with Spearman's and linear models.
Main Results:
- Heparin significantly affected VCM and TEG parameters, with differing impacts between unfractionated heparin (UFH) and low molecular weight heparin (LMWH).
- Significant discrepancies observed in clotting times and amplitudes between VCM and TEG under UFH and LMWH.
- VCM-MCF demonstrated a strong correlation with platelet count, superior to TEG-MA, particularly during UFH administration.
Conclusions:
- VCM and TEG exhibit distinct responses to heparin, necessitating careful interpretation in clinical practice.
- VCM-MCF provides a reliable assessment of platelet count, even in patients receiving UFH.
- Further research is warranted to fully elucidate the clinical utility of VCM compared to TEG.