Related Experiment Video
Updated: Jul 20, 2026

Quantification of the Immunosuppressant Tacrolimus on Dried Blood Spots Using LC-MS/MS
Published on: November 8, 2015
Evaluating the performance of the Roche FEN2 fentanyl immunoassay and its clinical implementation: The role of
Marlen Menlyadiev1, Raymond T Suhandynata1,2, Kyle Lund1
1Department of Pathology, Center for Advanced Laboratory Medicine, University of California, San Diego Health Systems, San Diego, CA, United States.
Introduction:
While laboratory-developed tests (LDTs) using liquid chromatography tandem mass spectrometry (LC-MS/MS) are widely employed to support the development of FDA-cleared drug immunoassays, their significance in the clinical implementation and evaluation of such assays is often overlooked. This paper reports on the important role of LC-MS/MS LDTs in demonstrating improved performance of the Roche FEN2 fentanyl immunoassay compared with the Thermo DRI fentanyl immunoassay.
Methods:
The FEN2 assay was implemented according to the manufacturer's instructions and its performance was compared to the existing DRI assay using LC-MS/MS as a reference. Clinical sensitivity and specificity were determined using 250 consecutive random patient specimens. Spiking experiments were conducted to determine cross-reactivity with 31 fentanyl analogs. Select DRI false-positive samples were analyzed by the FEN2 assay via time-of-flight mass spectrometry method (LC-QTOF).
Results:
The FEN2 assay showed improved clinical sensitivity compared to the DRI (98% vs 61%) in 250 consecutive patient samples due to its ability to detect norfentanyl. It also showed better clinical specificity by correctly classifying select DRI false-positive results. Upon implementation in clinical practice, the FEN2 resulted in a higher screening positivity rate than the DRI (17.3% vs 13.3%) and a greater LC-MS/MS confirmation rate of immunoassay-positive samples (96.8% vs 88.8%, respectively).
Conclusion:
The use of LC-MS/MS LDTs demonstrated that the FEN2 assay has greater clinical sensitivity and is less prone to false-positives than the DRI assay. These findings support the use of FEN2 in routine clinical practice and emphasize the role of mass spectrometry-based LDTs in clinical toxicology testing.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
10:13Author Spotlight: An Efficient Methodology to Confidently Differentiate and Characterize Fentanyl Analogs
Published on: November 8, 2024
Related Concept Videos
MALDI-TOF Mass Spectrometry
Rapid Identification of Pathogens
Automated Microbial Diagnostics