Experimental Analysis of Radiation Protection Offered by a Novel Exoskeleton-Based Radiation Protection System versus
Maria Katsarou1, Bruce Zwiebel2, Rajarshi Pal Chowdhury3
1Houston Methodist DeBakey Heart and Vascular Center, Houston Methodist Hospital, Houston, Texas; Section of Vascular Surgery, Fondazione Istituto di Ricovero e Cura a Carattere Scientifico Ca'Granda Ospedale Maggiore Policlinico, Milan, Italy.
Purpose:
To evaluate the radiation protection offered by an exoskeleton-based radiation protection system (Stemrad MD) and to compare it with that offered by conventional lead aprons.
Methods:
The experimental setup involved 2 anthropomorphic phantoms, an operator, a patient, and a C-arm as the x-ray radiation source. Thermoluminescent detectors were used to measure radiation doses to different radiosensitive body parts on the operator phantom both with the exoskeleton and a conventional lead apron at the left radial and right femoral positions. Detected radiation doses for the exoskeleton and lead apron for different body parts and positions were compared.
Results:
At the left radial position, the mean radiation dose (mGy) reduction by the exoskeleton compared with that by the lead apron was >90% for the left eye lens (0.22 ± 0.13 vs 5.18 ± 0.08; P < .0001), right eye lens (0.23 ± 0.13 vs 4.98 ± 0.10; P < .0001), left head (0.11 ± 0.16 vs 3.53 ± 0.07; P < .0001), right head (0.27 ± 0.09 vs 3.12 ± 0.10; P < .0001), and left brain (0.04 ± 0.08 vs 0.46 ± 0.07; P < .0001). At the right femoral position, radiation reduction was >90% for the left eye lens (0.14 ± 0.10 vs 4.16 ± 0.09; P < .0001), right eye lens (0.06 ± 0.08 vs 1.90 ± 0.11; P < .0001), left head (0.10 ± 0.08 vs 4.39 ± 0.08; P < .0001), left brain (0.03 ± 0.07 vs 1.44 ± 0.08; P < .0001), right brain (0.00 ± 0.14 vs 0.11 ± 0.13; P = .06), and thyroid (0.04 ± 0.07 vs 0.27 ± 0.09; P < .0001). Protection of the torso was equivalent to that offered by conventional lead aprons.
Conclusions:
The exoskeleton-based system provided superior radiation protection to the physician compared with that provided by conventional lead aprons. The effects are particularly impactful for the brain, eye lens, and head areas.