Assessment of Large Animal Vascular Dimensions for Intra-Aortic Device Research and Development: A Systematic Review
Gabriel Georges1, Thomas Couture2, Pierre Voisine1
1Cardiac Surgery Division, Quebec Heart and Lung Institute, QC, Canada.
Innovations (Philadelphia, Pa.)
|April 8, 2023
Summary
Large animal models like swine, sheep, and calves over 80 kg offer the best aortic diameter resemblance for cardiovascular device evaluation. Careful consideration of access sites, such as the common femoral artery, is crucial for smaller animal models.
Area of Science:
- Biomedical Engineering
- Comparative Anatomy
- Medical Device Development
Background:
- Pre-clinical evaluation of cardiovascular medical devices necessitates appropriate animal models.
- First-generation devices, including aortic endovascular prostheses and circulatory support systems, often present unique size and anatomical challenges.
- A significant data gap exists regarding the vascular dimensions of large animal models, hindering optimal model selection.
Approach:
- A comprehensive literature review was conducted on vascular dimensions in swine, ovine, and bovine models.
- Data were systematically collected from Embase and MEDLINE databases, focusing on the descending aorta and ilio-femoral arteries.
- Computed tomography angiography data from sheep and calves were integrated to augment existing literature findings.
Key Points:
- Vascular dimensions in swine, ovine, and bovine models were categorized by weight (40-60 kg, 61-80 kg, >80 kg).
- Swine, sheep, and calves exceeding 80 kg demonstrate the greatest similarity in aortic diameter to human anatomy.
- For smaller animal models, careful assessment of access site suitability, specifically the common femoral artery, is recommended for device implantation.
Conclusions:
- Large animal models, particularly swine, sheep, and calves weighing over 80 kg, are recommended for preclinical cardiovascular device testing due to anatomical similarities.
- The selection of appropriate animal models with relevant vascular dimensions is critical for the successful evaluation of novel cardiovascular devices.
- Further research into the vascular anatomy of large animals can improve the translational validity of pre-clinical studies.