Evaluating Performance of Circular Staplers Using Comparative Test Methods for Evidence-Based Surgery
Haley Strassner1, Alexander Caulk1, Nicholas Reher1
1Surgical Innovations, Medtronic plc, North Haven, CT, USA.
Surgical Innovation
|April 8, 2023
Summary
Triple-row staplers showed higher leak pressure and staple line perfusion compared to double-row staplers. Novel preclinical testing methods were developed to assess surgical stapling technology advancements.
Area of Science:
- Surgical innovation
- Biomedical engineering
- Medical device testing
Background:
- Anastomotic leak is a significant postoperative concern in circular stapler surgery.
- Variability in testing methods hinders the assessment and comparison of new surgical stapling technologies.
- Robust preclinical test methods are essential for advancing stapling device technology.
Purpose of the Study:
- To develop and validate novel, reproducible preclinical test methods for evaluating surgical staplers.
- To compare the performance of triple-row (TriEEA) and double-row (ECP) circular staplers using these new methods.
Main Methods:
- Quantified leak pressure, staple line perfusion, staple line security, and device removal force.
- In vivo testing for leak and perfusion; synthetic medium testing with Instron for removal force and security.
- Statistical analysis using unpaired student's t-test or Kruskal-Wallis test (P < .05).
Main Results:
- Triple-row staplers (TriEEA) exhibited 73% higher leak pressure than double-row staplers (ECP).
- TriEEA showed significantly higher staple line perfusion (148%) and required 78% less force for device removal.
- However, TriEEA staple line failure force was lower than ECP at specific pressures.
Conclusions:
- The study presents novel, reproducible methodologies for assessing surgical stapler performance.
- These methods enhance clinical relevance and facilitate direct comparisons between different stapling devices.
- The findings demonstrate the utility of the new test methods in evaluating key design features of triple-row versus double-row staplers.