Related Experiment Video
Updated: Aug 3, 2025

Author Spotlight: High-Quality Quantum Dot Nanobeads for Sensitive Fluorescent Lateral Flow Immunoassays
Published on: June 28, 2024
A fluorescent lateral flow immunoassay based on CdSe/CdS/ZnS quantum dots for sensitive detection of olaquindox in
Mei Hu1, Xiaofei Hu2, Guoqiang Wang3
1College of Food Science and Technology, Henan Agricultural University, Zhengzhou, Henan 450002, China.
Abstract:
A portable fluorescence immunosensor based on the CdSe/CdS/ZnS quantum dots (QDs) with multiple-shell structure was fabricated for the precise quantification of olaquindox (OLA). The QDs labeled anti-OLA antibody used as bioprobe played an important role in the design and preparation of a lateral flow test strip. Due to the strong fluorescent intensity of QDs, the sensitivity is greatly improved. The quantitative results were obtained using a fluorescent strip scan reader within 8 min, and the calculated limit of detection for OLA at 0.12 µg/kg, which was 2.7 times more sensitive than that of the conventional colloidal gold-based strips method. Acceptable recovery of 85.0%-95.5% was obtained by the spiked samples. This newly established QDs-based strip immunoassay method is suitable for the on-site detection and rapid initial screening of OLA in swine feedstuff, and is potentially applied for the detection of other veterinary drugs to ensure food safety.
More Related Videos
10:10Development of a Lateral Flow Immunochromatographic Strip for Rapid and Quantitative Detection of Small Molecule Compounds
Published on: November 13, 2021
06:18Multiplexed Isothermal Amplification Based Diagnostic Platform to Detect Zika, Chikungunya, and Dengue 1
Published on: March 13, 2018