Measuring the Complete-arch Distortion of an Optical Dental Impression
Published on: May 30, 2019
Predictability of intraoral scanner error for full-arch implant-supported rehabilitation
Francesco Zingari1,2,3, Matteo Meglioli1, Francesco Gallo3,4
1Center of Dental Medicine, Department of Medicine and Surgery, University of Parma, Via Gramsci 14, 43126, Parma, Italy.
Objectives:
The present study aimed to analyze the behaviors of three intraoral scanners (IOSs): evaluating the interdistance and axial inclination discrepancies in full-arch scans, predictable errors were searched.
Materials And Methods:
Six edentulous sample models with variable numbers of dental implants were used; reference data were obtained with a coordinate-measuring machine (CMM). Each IOS (i.e., Primescan, CS3600, and Trios3) performed 10 scans per model (180 total scans). The origin of each scan body was used as a reference point to measure interdistance lengths and axial inclinations. Precision and trueness of interdistance measurements and axial inclinations were evaluated to address error predictability. Bland-Altman analysis, followed by linear regression analysis and Friedman's test (plus Dunn's post hoc correction), was performed to evaluate the precision and trueness.
Results:
Regarding interdistance, Primescan showed the best precision (mean ± SD: 0.047 ± 0.020 mm), while Trios3 underestimated the reference value more than the others (p < 0.001) and had the worst performance (mean ± SD: -0.079 ± 0.048 mm). Concerning the inclination angle, Primescan and Trios3 tended to overestimate angle values, while CS3600 underestimated them. Primescan had fewer inclination angle outliers, but it tended to add 0.4-0.6° to the measurements.
Conclusions:
IOSs showed predictable errors: they tended to overestimate or underestimate linear measurements and axial inclinations of scan bodies, one added 0.4-0.6° to the angle inclination values. In particular, they showed heteroscedasticity, a behavior probably related to the software or the device itself.
Clinical Significance:
IOSs showed predictable errors that could affect clinical success. When performing a scan or choosing a scanner, clinicians should clearly know their behaviors.