Related Experiment Video
Updated: Aug 2, 2025

Continuous Venous-Arterial Doppler Ultrasound During a Preload Challenge
Published on: January 20, 2023
Automatic non-invasive blood pressure measurement in left ventricular assist device patients with a
Péter Nagy1, Balázs Sax2, Anna Kozák2
1Department of Measurement and Information Systems, Faculty of Electrical Engineering and Informatics, Budapest University of Technology and Economics, Budapest, Hungary.
Introduction:
Non-invasive measurement of blood pressure in patients with a left ventricular assist device (LVAD) is challenging due to the mechanical properties of these circulatory support devices. Keeping blood pressure in a target range is crucial to reduce the risk of LVAD complications. As current assessment methods require trained personnel resulting in infrequent monitoring, devices for home-based measurements are sorely needed.
Objectives:
In this paper, we report a measurement method and a custom-made monitoring device based on photoplethysmography (PPG) enabling automated measurement of the mean arterial pressure (MAP) of patients with LVAD.
Methods:
The method and the device were tested on 21 adult patients with LVAD, and the estimated MAP values were compared to MAP values measured simultaneously by a human expert applying the commonly used Doppler-based method.
Results:
Results of the proposed method showed highly significant correlation with Doppler-based MAP values (R = 0.85 for inflation, p < 0.001; R = 0.96 for deflation, p<0.001). The mean difference of the proposed method to the Doppler-based MAP values was 1.48 ± 5.43 mmHg for inflation and -0.19 ± 2.71 mmHg for deflation.
Conclusion:
The results demonstrate that the proposed method is a promising direction to achieve accurate, automated, non-invasive BP measurement, applicable in home health monitoring for patients with LVAD.
More Related Videos
03:40Point-of-Care Ultrasound for Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation Without Left Ventricular Venting
Published on: January 17, 2025
08:49Author Spotlight: Enhancing Graft Viability Assessment Through Quantitative Metrics and Innovative Reservoir Systems
Published on: August 2, 2024
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure in the Leg
Preparation: