Related Experiment Video
Updated: Aug 2, 2025

The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
A randomized controlled trial comparing three supraglottic airway devices used as a conduit to facilitate tracheal
Pascal Laferrière-Langlois1,2, Alexandre Dion3, Éric Guimond3
1Department of Anesthesiology, Faculty of Medicine and Health Sciences, Centre intégré universitaire de santé et services sociaux de l'Estrie-Centre hospitalier universitaire de Sherbrooke (CIUSSS de l'Estrie-CHUS), Sherbrooke, QC, Canada. pascal.laferriere-langlois@usherbrooke.ca.
Purpose:
Once difficult ventilation and intubation are declared, guidelines suggest the use of a supraglottic airway (SGA) as a rescue device to ventilate and, if oxygenation is restored, subsequently as an intubation conduit. Nevertheless, few trials have formally studied recent SGA devices in patients. Our objective was to compare the efficacy of three second-generation SGA devices as conduits for bronchoscopy-guided endotracheal intubation.
Methods:
In this prospective, single-blinded three-arm randomized controlled trial, patients with an American Society of Anesthesiologists Physical Status of I-III undergoing general anesthesia were randomized to bronchoscopy-guided endotracheal intubation using AuraGain™, Air-Q® Blocker, or i-gel® devices. We excluded patients with contraindications to an SGA or drugs and who were pregnant or had a neck, spine, or respiratory anomaly. The primary outcome was intubation time, measured from SGA circuit disconnection to CO2 measurement. Secondary outcomes included ease, time, and success of SGA insertion; success of intubation on first attempt; overall intubation success; number of attempts to intubate; ease of intubation; and ease of SGA removals.
Results:
One hundred and fifty patients were enrolled from March 2017 to January 2018. Median intubation times were similar across the three groups (Air-Q Blocker, 44 sec; AuraGain, 45 sec; i-gel, 36 sec; P = 0.08). The i-gel was faster to insert (i-gel: 10 sec; Air-Q Blocker, 16 sec; AuraGain, 16 sec; P < 0.001) and easier to insert (Air-Q Blocker vs i-gel, P = 0.001; AuraGain vs i-gel, P = 0.002). Success of SGA insertion, success of intubation, and number of attempts were similar. The Air-Q Blocker was easier to remove than the i-gel (P < 0.001).
Conclusion:
All three second-generation SGA devices performed similarly regarding intubation. Despite minor benefits of the i-gel, clinicians should select their SGA based on clinical experience.
Study Registration:
ClinicalTrials.gov (NCT02975466); registered on 29 November 2016.
More Related Videos
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Suctioning II: Procedure