Related Experiment Video
Updated: Aug 2, 2025

Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
Hybrid pedicle screw and modified cortical bone trajectory technique in transforaminal lumbar interbody fusion at
Alafate Kahaer1, Rui Zhang1, Yixi Wang2
1Department of Spine Surgery, The First Affiliated Hospital of Xinjiang Medical University, 137 Liyushan Road, Urumqi, China.
Background:
Investigate the biomechanical properties of the hybrid fixation technique with bilateral pedicle screw (BPS) and bilateral modified cortical bone trajectory screw (BMCS) in L4-L5 transforaminal lumbar interbody fusion (TLIF).
Methods:
Three finite element (FE) models of the L1-S1 lumbar spine were established according to the three human cadaveric lumbar specimens. BPS-BMCS (BPS at L4 and BMCS at L5), BMCS-BPS (BMCS at L4 and BPS at L5), BPS-BPS (BPS at L4 and L5), and BMCS-BMCS (BMCS at L4 and L5) were implanted into the L4-L5 segment of each FE model. The range of motion (ROM) of the L4-L5 segment, von Mises stress of the fixation, intervertebral cage, and rod were compared under a 400-N compressive load with 7.5 Nm moments in flexion, extension, bending, and rotation.
Results:
BPS-BMCS technique has the lowest ROM in extension and rotation, and BMCS-BMCS technique has the lowest ROM in flexion and lateral bending. The BMCS-BMCS technique showed maximal cage stress in flexion and lateral bending, and the BPS-BPS technique in extension and rotation. Compared to the BPS-BPS and BMCS-BMCS technique, BPS-BMCS technique presented a lower risk of screw breakage and BMCS-BPS technique presented a lower risk of rod breakage.
Conclusion:
The results of this study support that the use of the BPS-BMCS and BMCS-BPS techniques in TLIF surgery for offering the superior stability and a lower risk of cage subsidence and instrument-related complication.