[Preliminary survey report on the clinical validation of in-use electronic sphygmomanometers in China]
1Peking University First Hospital, Beijing 100034, China Peking University Clinical Research Institute, Beijing 100091, China.
Abstract:
Objective: To explore the percentage of in-use electronic sphygmomanometers independently validated clinically in China. Methods: We conducted a cross-sectional survey and Beijing, Shenzhen, Shijiazhuang, Datong, and Shihezi were selected according to the geographical location and economic level. In each site, one tertiary hospital, two community health centers, and 20 families with electronic sphygmomanometers in use were chosen. The information of electronic sphygmomanometers including brand, model, manufacturer and production date were obtained by the trained staff. Ten electronic sphygmomanometers from each hospital, five electronic sphygmomanometers from each community health center, and one electronic sphygmomanometer from each family were surveyed, and the user's subjective judgment results and judgment basis on the accuracy of the electronic sphygmomanometer measurement were collected. We searched six registration websites (Medaval, Stride BP, dabl Educational Trust, British and Irish Hypertension Society, American Medical Association and Hypertension Canada) and two research databases (PubMed and CNKI) for the clinical validation status of each electronic sphygmomanometer. Results: A total of 200 electronic sphygmomanometers were investigated in this study, of which only 29.0% (58/200) passed independent clinical validation. When stratified by users, the percentage of being clinical validated was 46.0% (23/50) for electronic sphygmomanometers in hospitals, 42.0% (21/50) for those in community health centers and 14.0% (14/100) for those in home use, respectively, and the proportions between the three groups were significantly difference (P<0.001). Doctors in tertiary hospitals and community health service centers judged the accuracy of electronic sphygmomanometers mainly on the basis of "regular correction" (41.0% (41/100)) and "comparison with other electronic sphygmomanometers" (20.0% (20/100)), while among home users, 41.0% (41/100) were not clear about the accuracy of electronic sphygmomanometers, and 40.0% (40/100) made the judgment by "comparison with the devices in hospitals". Conclusion: The clinical validation of in-use electronic sphygmomanometers in China is low. Most of users, including healthcare professionals, are not aware of clinical validation of electronic sphygmomanometers.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure