Related Experiment Video
Updated: Aug 2, 2025

Whole-body PET/MRI of Pediatric Patients: The Details That Matter
Published on: December 19, 2017
Pediatric computed tomography doses in Germany from 2016 to 2018 based on large-scale data collection
Daniel Wollschläger1, Andreas Jahnen2, Johannes Hermen2
1Institute of Medical Biostatistics, Epidemiology and Informatics, University Medical Center of the Johannes Gutenberg University Mainz, Mainz, Germany.
Pediatric computed tomography (CT) scans show decreasing radiation doses, supporting lower dose reference levels (DRL). This study collected CT scan data to update DRLs for safer pediatric imaging.
Area of Science:
- Medical Imaging
- Radiological Physics
- Pediatric Radiology
Background:
- Pediatric CT scans carry a small risk of leukemia or brain tumors.
- Optimizing radiation doses in pediatric CT is crucial.
- Dose reference levels (DRL) help reduce collective radiation exposure.
Purpose of the Study:
- To collect dosimetric data for pediatric CT examinations.
- To support the adaptation of current DRLs based on evolving clinical practices.
- To assess if current DRLs need adjustment due to technological advancements and protocol optimization.
Main Methods:
- Retrospective collection of dosimetric data and technical parameters.
- Data gathered from Picture Archiving and Communication Systems (PACS), Dose Management Systems (DMS), and Radiological Information Systems (RIS).
- Inclusion of common pediatric CT examinations across 17 institutions from 2016-2018.
Main Results:
- Data from 7746 CT series in patients under 18 years were analyzed.
- Age-stratified parameter distributions were generally lower than pre-2010 data.
- Most third quartile dose values were below the German DRL at the time of the survey.
Conclusions:
- Large-scale data collection via PACS, DMS, and RIS is feasible but requires high data quality.
- Data validation through expert knowledge or questionnaires is recommended.
- Observed practices suggest a reasonable basis for lowering some DRLs in Germany for pediatric CT.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...