Related Experiment Video
Updated: Aug 2, 2025

Fully Endoscopic Mitral Valve Repair with Percutaneous Cannulation of Groin Vessels
Published on: May 26, 2023
Cutting balloon for femoral arterial and venus obstructions due to suture-based closure devices: Case series
Ilan Merdler1, Nelson L Bernardo1, Itsik Ben-Dor1
1Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, District of Columbia, USA.
Abstract:
Suture-based vascular closure devices have been shown to be effective in hemostasis for procedures with vascular access. However, iatrogenic vascular occlusion may occur. The cutting balloon (CB) is a noncompliant balloon wrapped with 3-4 microsurgical blades that are intended to modify vascular lesions, but it may also be utilized to cut and release endovascular sutures. We report two cases in which the CB was employed as a bailout strategy to alleviate suture-related vascular occlusion after transcatheter aortic valve replacement. The CB can be effectively utilized to resolve suture-related vascular occlusion.