Related Experiment Video
Updated: Aug 2, 2025

Three-dimensional Optical-resolution Photoacoustic Microscopy
Published on: May 3, 2011
Single-shot photoacoustic imaging with single-element transducer through a spatiotemporal encoder
1California Institute of Technology, Caltech Optical Imaging Laboratory, Andrew and Peggy Cherng Department of Medical Engineering, Pasadena, California, United States.
Significance:
Current photoacoustic (PA) imaging modalities typically require either serial detection with a single-element transducer or parallel detections with an ultrasonic array, indicating a dilemma between system cost and imaging throughput. PA topography through ergodic relay (PATER) was recently developed to address this bottleneck. However, PATER requires object-specific calibration due to varied boundary condition and must be recalibrated through pointwise scanning for each object before measurements, which is time-consuming and severely limits practical application.
Aim:
We aim to develop a new single-shot PA imaging technique that only requires a one-time calibration for imaging different objects using a single-element transducer.
Approach:
We develop an imaging method, PA imaging through a spatiotemporal encoder (PAISE), to address the above issue. The spatial information is effectively coded into unique temporal features by the spatiotemporal encoder, which allows for compressive image reconstruction. An ultrasonic waveguide is proposed as a critical element to guide the PA waves from the object into the prism, which effectively accounts for the varied boundary condition of different objects. We further add irregular-shaped edges on the prism to introduce randomized internal reflections and further facilitate the scrambling of acoustic waves.
Results:
The proposed technique is validated through comprehensive numerical simulations and experiments, and it is demonstrated that PAISE can successfully overcome the changed boundary condition and can image different samples given a single calibration.
Conclusions:
The proposed PAISE technique is capable of single-shot widefield PA imaging with a single-element transducer and does not require sample-specific calibration, which successfully overcomes the major limitation of previous PATER technology.
More Related Videos
07:14Dual Raster-Scanning Photoacoustic Small-Animal Imager for Vascular Visualization
Published on: July 15, 2020
09:56Universal Hand-held Three-dimensional Optoacoustic Imaging Probe for Deep Tissue Human Angiography and Functional Preclinical Studies in Real Time
Published on: November 4, 2014