Related Experiment Video
Updated: Aug 2, 2025

Application of Design Aspects in Uniaxial Loading Machine Development
Published on: September 19, 2018
A structural analysis of the rail unit of an indoor assistive mobility system
Je Nam Kim1, Mi Yeon Shin2, You Sung Seo3
1New Technology Convergence Team, R&BD Division, CAMTIC Advanced Mechatronics Technology Institute for Commercialization, Jeonju-si, Korea.
Background:
Individuals with gait disturbances, such as that post-stroke, are discharged home to undergo outpatient rehabilitation. Rehabilitation in the community is not as effective as that in hospital, due to long travel times and short program duration.
Objective:
This study analyzed rail unit structure, with the aim of assisting home indoor assistive mobility system (HIAMS) development, allowing patients to undergo gait-related rehabilitation training at home.
Methods:
The HIAMS consists of a mobile rail running around the whole room, a turn-table for movement between rails, and a weight-supporting component. Structural analysis was performed using the Abaqus/CAE solution (Version 6.14, Dassault systems, Inc.) to verify device safety, according to the load applied to the rail and turn-table units. The load was applied vertically at 150 kg to reflect the weight of potential users.
Results:
Structural analysis was performed on the weight-supporting components, which was consist of turn-table case, bearing components (center, left), connective bracket and rail rollers. The safety factors of each components were estimated as 1.31, 5.39 (bearing, center), 8.45 (bearing, left), 1.43 and 3.61 in sequence.
Conclusion:
We demonstrated a safety factor of ⩾ 1.3 for the key system units, suggesting this technology is safe for use in the home rehabilitation training of individuals with gait impairment post-ICU stay.
Related Concept Videos
Relative Motion Analysis - Acceleration
Machines: Problem Solving II