Influence on voxel-based dosimetry: noise effect on absorbed dose dosimetry at single time-point versus sequential
Uysha de S Fonda1, André L A Leitão2, Marcia M D P Paiva3
1Departmento de Radiologia e Oncologia da Faculdade de Medicina, Universidade de Sao Paulo, Sao Paulo.
Objective:
The purpose of this study was to evaluate how statistical fluctuation in single-photon emission computed tomography (SPECT) images propagate to absorbed dose maps.
Methods:
SPECT/computed tomography (CT) images of iodine-131 filled phantoms, using different acquisition and processing protocols, were evaluated using STRATOS software to assess the absorbed dose distribution at the voxel level. Absorbed dose values and coefficient of variation (COV) were analyzed for dosimetry based on single time-point SPECT images and time-integrated activities of SPECT sequences with low and high counts.
Results:
Considering dosimetry based on a single time-point, the mean absorbed dose was not significantly affected by total counts or reconstruction parameters, but the uniformity of the absorbed dose maps had an almost linear correlation with SPECT noise. When high- and low-count SPECT sequences were used to generate an absorbed dose map, the absorbed dose COV for each of the temporal sequences was slightly lower than the absorbed dose COV based on the single SPECT image with the highest count included in the sequence.
Conclusion:
The impact of changes in SPECT counts and reconstruction parameters is almost linear when dosimetry is based on isolated SPECT images, but less pronounced when dosimetry is based on sequential SPECTs.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Drug Concentration Versus Time Correlation
Two pivotal parameters are the minimum effective concentration (MEC) and the minimum toxic concentration (MTC). The MEC is the...
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Biological Effects of Radiation