Comparison of the Hologic Genius Digital Diagnostics System with the ThinPrep Imaging System-A retrospective
Hans Ikenberg1, Sabrina Lieder2, André Ahr1,3
1MVZ CytoMol, Frankfurt, Germany.
Background:
Digital cytology (DC) with artificial intelligence (AI) is a new approach. The authors compared DC with liquid-based cytology (LBC) using computer assistance (CAS) in a retrospective, noninterventional study.
Methods:
In total, 1994 ThinPrep LBC slides (Hologic), which were previously analyzed in 2020 using an imaging system with CAS in routine cotesting for cytology/human papillomavirus, were reviewed in a blinded mode using the Genius Digital Diagnostics System (Hologic). In 555 cases, a histology result was available. The slides were digitally scanned (volumetric scan) at 14 levels integrated into one. AI algorithms were used to present a gallery of six tiles each (containing objects of interest) in five categories. Six additional tile rows were available, from which the diagnoses were made. All cases with a mismatch between DC and imaging system results were reviewed by an additional cytopathologist.
Results:
In 86.56% of cases, a complete match between both systems was observed using the same cytology categories. When also considering the histology results, the match was 90.37%. In addition, when a cytology follow-up and/or a retrospective review was applied, the match reached 97.34%. In only 0.65% of cases was a major discrepancy observed (two grades of cytology or a low-grade squamous intraepithelial lesion/high-grade squamous intraepithelial lesion [LSIL/HSIL] shift), and none were identified by DC. Significantly more cases of higher severity (atypical squamous cells cannot exclude high grade [ASC-H], high-grade squamous intraepithelial lesion [HSIL]) were identified with DC, and its negative predictive value was higher. The screening time was significantly shorter with DC.
Conclusions:
With the Genius system for DC, the sensitivity for HSIL+/ASC-H and the specificity for LSIL and HSIL were superior to LBC and CAS. Screening time was significantly lower.
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...