Full-Field Optical Coherence Microscopy for Histology-Like Analysis of Stromal Features in Corneal Grafts
Published on: October 21, 2022
1.7K
Comparison of central corneal thickness measurements using three different imaging devices
1Medicana International Hospital, Department of Ophthalmology, Samsun, Turkey.
Journal Francais D'Ophtalmologie
|April 19, 2023
Summary
Central corneal thickness (CCT) measurements varied between devices. AL-Scan underestimated CCT compared to ultrasonic pachymetry and OCT, indicating devices are not interchangeable for clinical use.
Area of Science:
- Ophthalmology
- Biomedical Optics
- Medical Device Technology
Background:
- Accurate central corneal thickness (CCT) measurement is crucial for ophthalmic diagnostics and surgical planning.
- Various devices exist for CCT assessment, but their comparative performance and agreement require thorough evaluation.
Purpose of the Study:
- To compare central corneal thickness (CCT) values obtained from three distinct measurement devices.
- To evaluate the inter-device agreement for CCT measurements in healthy eyes.
Main Methods:
- Retrospective analysis of CCT data from 120 eyes of 60 healthy individuals.
- Measurements were performed using an optical biometer (AL-Scan), spectral-domain optical coherence tomography (SD-OCT), and ultrasonic pachymetry (UP).
- Bland-Altman analysis was employed to assess the agreement between the measurement methods.
Main Results:
- Mean CCT values were 532.4μm (AL-Scan), 549μm (UP), and 547μm (SD-OCT).
- Significant differences were found between AL-Scan and both UP (17.15±8.42μm) and SD-OCT (15.30±9.52μm).
- A close correlation was observed between all three CCT measurement methods, with minimal difference between UP and SD-OCT (1.85±8.78μm).
Conclusions:
- While all devices showed good correlation, AL-Scan significantly underestimated CCT compared to UP and SD-OCT.
- Clinicians must recognize that different devices yield varying CCT results.
- To ensure consistent patient management, especially pre-refractive surgery, CCT assessment should utilize the same device consistently.
Keywords:
Anterior segmentCorneaCorneal thickness measurementCornéeMesure de l’épaisseur de la cornéePachymetryPachymétrieSegment antérieur de l’œilUltrasonographyÉchographie