Anterior eye parameters and lens thickness measured by an intraoperative OCT and a swept-source OCT: comparison of
Michael Müller1, Charlotte Wortmann2, Julia Paul1
1Department of Ophthalmology, Goethe-University, Frankfurt am Main, Germany.
Purpose:
To evaluate if anterior chamber depth (ACD) and lens thickness (LT) measured by two different devices are affected by different eye lengths.
Methods:
ACD and LT of 251 eyes (44 hyperopic, 60 myopic, 147 emmetropic) of 173 patients received with an iOCT-guided femtosecond laser-assisted lens surgery (FLACS) and the IOL Master 700 were compared.
Results:
ACD measured with the IOL Master 700 was -0.026 ± 0.125 mm smaller (p = 0.001) than that with the iOCT for all eye-groups (hyperopic: p = 0.601, emmetropic: p = 0.003; myopic: p = 0.094). However, differences in all groups were not clinically relevant. LT measurements (all eyes: -0.0642 ± 0.0504 mm) shows a statistically significant difference in all evaluated groups (p < 0.001). Only myopic eyes showed a clinically relevant difference in LT.
Conclusion:
The two devices show no clinically relevant differences in the eye-length groups (myopic, emmetropic, and hyperopic) for all ACD measurements. LT data shows a clinically relevant difference only for the group of myopic eyes.