Specimen-specific finite element representations of implanted hip capsules
Related Experiment Video
Updated: Aug 2, 2025

Imaging of the Microstructural Failure Mechanism in the Human Hip
Published on: September 29, 2023
Ahilan Anantha Krishnan1, Casey A Myers1, Michael Scinto1
1Center for Orthopaedic Biomechanics, University of Denver, Denver, CO, USA.
Abstract:
The hip capsule is a ligamentous structure that contributes to hip stability. This article developed specimen-specific finite element models that replicated internal-external (I-E) laxity for ten implanted hip capsules. Capsule properties were calibrated to minimize root mean square error (RMSE) between model and experimental torques. RMSE across specimens was 1.02 ± 0.21 Nm for I-E laxity and 0.78 ± 0.33 Nm and 1.10 ± 0.48 Nm during anterior and posterior dislocation, respectively. RMSE for the same models with average capsule properties was 2.39 ± 0.68 Nm. Specimen-specific models demonstrated the importance of capsule tensioning in hip stability and have relevance for surgical planning and evaluation of implant designs.