Safety and effectiveness comparison of a new cohesive ophthalmic viscosurgical device
Mark Packer1, Mitchell Shultz, James Loden
1From the Mark Packer MD Consulting, Inc., Boulder, Colorado (Packer); Shultz Chang Vision, Northridge, California (Shultz); Loden Vision Center, Nashville, Tennessee (Loden); Bausch & Lomb, Inc., Bridgewater, New Jersey (Lau).
Purpose:
To evaluate the safety and effectiveness of a new cohesive ophthalmic viscosurgical device (OVD) (StableVisc) compared with a marketed cohesive OVD (ProVisc) in patients undergoing cataract surgery.
Setting:
22 sites in the United States.
Design:
Prospective multicenter controlled double-masked and randomized 1:1 (StableVisc:ProVisc; stratified by site, age group, and cataract severity).
Methods:
Adults (≥45 years) with age-related noncomplicated cataract considered amenable to treatment with standard phacoemulsification cataract extraction and intraocular lens implantation were included. Patients were randomized to receive either StableVisc or ProVisc during standard cataract surgery. Postoperative visits occurred at 6 hours, 24 hours, 7 days, 1 month, and 3 months. The primary effectiveness outcome was the change in endothelial cell density (ECD) from baseline to 3 months. The primary safety endpoint was the proportion of patients who experienced at least 1 intraocular pressure (IOP) measurement ≥30 mm Hg at any follow-up visit. Noninferiority between the devices was tested. Inflammation and adverse events were evaluated.
Results:
390 patients were randomized; 187 patients with StableVisc and 193 patients with ProVisc completed the study. StableVisc was noninferior to ProVisc in mean ECD loss from baseline to 3 months (17.5% and 16.9%, respectively). StableVisc was noninferior to ProVisc in the proportion of patients with postoperative IOP ≥30 mm Hg at any follow-up visit (5.2% and 8.2%, respectively).
Conclusions:
The StableVisc cohesive OVD, which provides both mechanical and chemical protection, was safe and effective when used in cataract surgery and provides surgeons with a new cohesive OVD.
More Related Videos
07:27Anterior Segment Organ Culture Platform for Tracking Open Globe Injuries and Therapeutic Performance
Published on: August 25, 2021
08:05Occlusion of the Great and Small Saphenous Vein Using Copolymeric Glue Based on N-Butyl Cyanoacrylate and Methacryloxy Sulfolane
Published on: December 9, 2022
Related Concept Videos
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Endoscopic Procedures III: Video Capsule Endoscopy