Ultra-high frequency lung ultrasound in preterm neonates: a test validation study on interpretation agreement and
Victor Sartorius1, Barbara Loi2,3, Laura Vivalda2
1Division of Paediatric and Neonatal Critical Care, Hôpital Antoine-Béclère, Clamart, France victor.sartorius@aphp.fr.
Objective:
To verify if increasing frequency, through the use of ultra-high frequency transducers, has an impact on lung ultrasound pattern recognition.
Design:
Test validation study.
Setting:
Tertiary academic referral neonatal intensive care unit.
Patients:
Neonates admitted with respiratory distress signs.
Interventions:
Lung ultrasound performed with four micro-linear probes (10, 15, 20 and 22 MHz), in random order. Anonymised images (600 dpi) were randomly included in a pictorial database: physicians with different lung ultrasound experience (beginners (n=7), competents (n=6), experts (n=5)) blindly assessed it. Conformity and reliability of interpretation were analysed using intraclass correlation coefficient (ICC), area under the curve (AUC) of the multi-class ROC analysis, correlation and multivariate linear regressions (adjusting for frequency, expertise and their interaction).
Outcome Measures:
A (0-3) score based on classical lung ultrasound semiology was given to each image as done in the clinical routine.
Results:
ICC (0.902 (95% CI: 0.862 to 0.936), p<0.001) and AUC (0.948, p<0.001) on the whole pictorial database (48 images acquired on 12 neonates), and irrespective of the frequency and physicians' expertise, were excellent. Physicians detected more B-lines with increasing frequency: there was a positive correlation between score and frequency (ρ=0.117, p=0.001); multivariate analysis confirmed the score to be higher using 22 MHz-probes (β=0.36 (0.02-0.7), p=0.041).
Conclusion:
Overall conformity and reliability of interpretations of lung ultrasound patterns were excellent. There were differences in the identification of the B-patterns and severe B-patterns as increasing probe frequency is associated with higher score given to these patterns.
More Related Videos
05:50International Expert Consensus and Recommendations for Neonatal Pneumothorax Ultrasound Diagnosis and Ultrasound-guided Thoracentesis Procedure
Published on: March 12, 2020
11:27A Modified Sonographic Algorithm for Image Acquisition in Life-Threatening Emergencies in the Critically Ill Newborn
Published on: April 7, 2023
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...