Optimization design of a permanent magnet used for a low field (0.2 T) movable MRI system
Shufeng Wei1, Zhao Wei1, Zheng Wang1
1Department of Electromagnetic Detection and Imaging Research, Institute of Electrical Engineering, Chinese Academy of Sciences, Zhongguancun Beiertiao NO.6, Beijing, 100190, China.
Objective:
To design a lightweight permanent magnet for a lowfield movable head imaging MRI system.
Materials And Methods:
To reduce the weight of the magnet, the pole pieces, anti-eddy current plates, and shimming rings were removed, and the distance between the two vertical yokes was shortened as much as possible. To compensate for the magnetic field deformation caused by the shortened distance between two vertical iron yokes, two side magnetic poles were added to the vertical yokes. The magnetic field distributions in magnetic poles, the iron yoke, and the spherical imaging region were simulated. Phantom and in vivo head imaging were conducted with a lowfield movable MRI prototype scanner equipped with the proposed permanent magnet.
Results:
A permanent magnet with a center field of 0.19815 T, a homogeneity of 46 ppm over the 20 cm spherical imaging region, and a weight of 654 kg have been achieved. Acceptable images of a phantom and a human brain have been acquired with the prototype MRI scanner.
Discussion:
The proposed permanent magnet design significantly reduces the magnet's weight compared with the conventional magnet structure and shows promise in promoting the development of lowfield compact MRI systems.
Related Concept Videos
Ferromagnetism
Magnetic Resonance Imaging
Magnetic Field of a Solenoid
Consider a solenoid with 100 turns wrapped around a cylinder of...
Paramagnetism
Magnetism
An individual magnetic pole cannot be isolated. No matter how small, every piece of a magnet contains a north pole and a south...
Magnetic Susceptibility and Permeability
When diamagnetic materials are placed under an external magnetic field, the moments opposite to the field are induced. Hence, the susceptibility for diamagnets has a minimal negative value of 10-5–10-6. Since...