Related Experiment Video
Updated: Aug 2, 2025

Diffusion Imaging in the Rat Cervical Spinal Cord
Published on: April 7, 2015
[Evaluation of the Effectiveness of a Neck Fixation Device for Improving Image Quality in Diffusion-weighted
Naotaka Kaminaga1,2, Tomoya Kobayashi3, Yoshiyuki Ishimori2
1Department of Radiology, Ibarakihigashi National Hospital, National Hospital Organization.
Abstract:
The effectiveness of a neck fixation device to improve the image quality of DWIBS was investigated. Healthy volunteers were examined while chewing with and without a neck fixation device using a 3-T MRI system. Distance of mandibular movement was measured using true-fast imaging of steady-state precession (true FISP). Signal-to-noise ratio (SNR) and apparent diffusion coefficient (ADC) of DWIBS were measured. Image quality of DWIBS was scored by visual evaluation. These values were compared with and without a neck fixation device. Regarding results, the mandibular displacement and ADC were decreased, and the SNR and visual score were increased by the use of the fixation device. There is a significant difference between with and without a neck fixation device in each measurement. The technique using a neck fixation device helps improve image quality of DWIBS in the head and neck region.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Magnetic Resonance Imaging
Imaging Studies III: Computed Tomography