Related Experiment Video
Updated: Aug 2, 2025

Detection of Architectural Distortion in Prior Mammograms via Analysis of Oriented Patterns
Published on: August 30, 2013
To what extent are objectively measured mammographic imaging techniques associated with compression outcomes
Sue M Hudson1, Louise S Wilkinson2, Bianca L De Stavola
1Department of Non-Communicable Disease Epidemiology, London School of Hygiene and Tropical Medicine, Keppel, London, UK.
Objective:
To describe the association between objectively measurable imaging techniques and the resulting compression thickness and dose.
Methods:
The study included 80,495 routine screens from the South-West London Breast Screening Service between March 2013 and July 2017. Average compression force, paddle tilt and dose were calculated. The Volpara® DensityTM algorithm was used to estimate pressure, breast volume and density.Linear regression models, using generalized estimating equations (GEEs) to account for clustering by practitioner, assessed the strength of the associations between the imaging compression outcomes, (thickness, dose) and imaging techniques (force, pressure and paddle tilt), adjusting for the subject's characteristics (age, ethnicity, breast volume and percent mammographic density).
Results:
Fully adjusted linear regression models showed that compression thickness decreased by ~1 mm (~2% of mean thickness) for every 1daN increase in force and decreased by ~0.8 mm with an increase of 1 kPa of pressure (at median pressure). Increasing pressure above 15 kPa resulted in minimal reduction in thickness. Dose increased with increased force but decreased by ~1% of mean dose with every increase in 1 kPa of pressure. For 1o increase in paddle tilt, the compression thickness increased by ~1.5 mm (~2.5%) and dose increased by ~2.5%, (Pt <0.001 in all cases).
Conclusion:
Differences in imaging technique are associated with imaging outcome measures (thickness and dose). A better understanding of the association between objective image acquisition parameters and tumour conspicuity could lead to clearer guidelines for practitioners.
Advances In Knowledge:
Increased paddle tilt is associated with increased compression thickness and increased dose after adjustment for breast volume and force applied.
More Related Videos
15:48Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
08:32Using Computer-based Image Analysis to Improve Quantification of Lung Metastasis in the 4T1 Breast Cancer Model
Published on: October 2, 2020
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Ultrasonography
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging