Stephen D Cook1, Laura P Patron1, Carlos J Lavernia2
1Fellowship of Orthopaedic Researchers, Metairie, Louisiana.
Background:
Fracture of contemporary femoral stems is a rare occurrence in total hip arthroplasty. A knowledge gap remains regarding manufacturing, patient, and surgeon factors that may contribute to the increased risk of this complication.
Methods:
We analyzed 13 contemporary fractured porous-coated femoral stems of various designs to determine cause and contributing factors of mechanical failure. Cases included 12 men and 1 woman who had an average age at index surgery of 53 years (range, 34 to 76 years). There were 10 of 13 patients who had a body mass index more than 30 (obese); 3 of the 10 had a body mass index more than 40. The mean time to fracture was 7.6 years (range, 7 months to 12 years).
Results:
There were 4 titanium alloy stems that fractured an average of 3.6 years postrevision surgery for head/cup exchange and had associated iatrogenic mechanical and electrocautery damage to the femoral neck at fracture initiation sites. There were 6 modular stems that failed at the stem-sleeve or stem-neck interfaces with evidence of fretting corrosion. For 2 stem-neck fractures, mismatched head/stem combinations from different manufacturers resulted in untested mechanical offsets and loading. There were 2 proximal neck fractures and 1 mid-shaft fracture of coated cobalt-chromium alloy stems that occurred in 3 obese men. The neck fractures (10 to 12 years) were well-fixed stems. Lack of proximal fixation contributed to the mid-shaft fracture (7 months).
Conclusion:
While rare, femoral stem fractures pose catastrophic outcomes in primary and revision total hip arthroplasty. Manufacturing, patient, and surgical factors contributing to stem failures were identified, including patient obesity, heat-treatment reduction of mechanical properties, iatrogenic implant damage, and mixing of different vendor stems and heads.
Related Concept Videos
Bones of the Lower Limb: Femur and Patella
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...