Simultaneous multislice imaging with slice-specific z-shim
Jürgen Finsterbusch1, Ying Chu1
1Department of Systems Neuroscience, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
Magnetic Resonance in Medicine
|April 24, 2023
Summary
Slice-specific z-shimming improves simultaneous multislice (SMS) imaging by minimizing signal loss in accelerated T2*-weighted scans. This technique enhances image quality for functional neuroimaging, particularly in the spinal cord.
Area of Science:
- Magnetic Resonance Imaging (MRI)
- Neuroimaging
- Pulse Sequence Design
Background:
- Simultaneous multislice (SMS) imaging accelerates MRI acquisition by exciting multiple slices at once.
- Slice-accelerated T2*-weighted imaging, crucial for functional neuroimaging, suffers from signal loss due to magnetic field inhomogeneities.
- Conventional z-shimming methods are not optimal for SMS due to the need for slice-specific magnetic field corrections.
Purpose of the Study:
- To implement and evaluate slice-specific z-shimming for simultaneous multislice (SMS) imaging.
- To minimize signal losses in slice-accelerated T2*-weighted acquisitions, specifically for spinal cord functional neuroimaging.
Main Methods:
- Developed a novel method using RF envelope shifts to achieve slice-specific z-shimming in SMS imaging.
- Performed echo-planar imaging (EPI) with 2-fold SMS acceleration on a 3T MRI system.
- Compared image quality and signal-to-noise ratio (SNR) across three conditions: no z-shim, conventional average z-shim, and slice-specific z-shim.
Main Results:
- Phantom experiments confirmed the equivalence of envelope shift and conventional z-shimming for non-SMS excitations.
- The "mixed" z-shim approach, combining envelope shifts for slice differences and conventional methods for the mean, yielded the best image quality in SMS.
- This optimized approach improved signal amplitude and temporal SNR in phantoms and in vivo compared to non-slice-specific methods, with minimal TE differences.
Conclusions:
- RF envelope shifts enable slice-specific z-shimming in SMS imaging.
- This technique has the potential to significantly improve slice-accelerated functional neuroimaging, particularly for the spinal cord.