Related Experiment Video
Updated: Aug 1, 2025

Rapid Point-of-Care Assay of Enoxaparin Anticoagulant Efficacy in Whole Blood
Published on: October 12, 2012
Optimization of laboratory diagnosis of heparin-induced thrombocytopenia using HemosIL-AcuStar-HIT-IgG assay
Catherine M Tucker1, Ruben Rhoades2, Ruchika Sharma3
1Department of Pathology, Anatomy, and Cell Biology, Thomas Jefferson University, Philadelphia, PA, US.
Objective:
The aim of this study was to determine an optimal cutoff value for the newly available HemosIL-AcuStar-HIT-IgG assay (AcuStar) for the diagnosis of heparin-induced thrombocytopenia (HIT).
Method:
We evaluated the performance of AcuStar using serotonin release assay (SRA) as the gold standard and incorporated 4T score calculation in a cohort of suspected HIT cases. Statistical analysis was performed to determine optimal cutoff value for the diagnosis of HIT.
Result:
A diagnosis of HIT can be excluded with a platelet factor 4 (PF4) value of <0.4 U/mL by AcuStar and 4T score in the low-risk category (≤3). All other cases will require confirmation with a functional test.
Conclusion:
Our study resulted in the implementation of a diagnostic algorithm for laboratory diagnosis of HIT, which incorporates pretest calculation of 4T score and AcuStar as a screening test, with reflex confirmation by SRA. This new algorithm resulted in extended hours of test availability and a more rapid turnaround time in reporting PF4 results.