Evaluation and Comparison of Stress in Divergent and Convergent Collar Designs of Implants With Different Bone
Gracy Panmei1, Arunoday Kumar2, Suchetana Basak2
1Prosthodontics and Crown & Bridge, Private Practice, Imphal, IND.
Background:
The failure and the success rate of an implant depends on biomechanical factors, esthetics and painless sterile implant surgery conditions, out of which stresses applied to the bone and its surrounding, bone-implant interface, material characteristics of the implant used and the strength of the bone and its surrounding are the important factors. This study aimed to evaluate the stress distribution of divergent collar design (DCD) and convergent collar design (CCD) implants placing them in four different densities of the bone (D1, D2, D3 and D4). The evaluation of the stress distribution of DCD and CCD was performed using the 3D finite element method (FEM), by placing them in four different bone densities. In addition to this, a comparison of the effect of DCD and CCD in terms of stress distribution in the bone was also done.
Materials And Methods:
The software used to process the geometric characteristics of the missing first molar in the mandibular section were Ansys, version 19.2, CATIA, version 5, and Solidworks (Dassault Systèmes). Using these software, three models were designed and successfully restored using an all-ceramic crown implant. The first model was a geometric model of the first molar mandibular bone section, the second model was a cylindrical implant (4x10 mm) with a DCD and CCD, and the third model had titanium alloy (Ti-6Al-4V) properties incorporated into the implant.
Results:
The D1 bone model showed the lowest stress concentration compared to D2, D3, D4. The DCD showed the lower stress and strain concentrations as compared to the CCD in the contiguous crestal bone in all the densities of the bone in both vertical and lateral or oblique loadings. The DCD with the D1 bone showed the least stress concentration around the crestal bone region. The results of this study also showed that the maximum von Mises stress was observed in the crestal region or the neck of the implant for both the convergent and divergent collar implant designs in all the four densities of the bone.
Conclusion:
Before a patient trial of a new implant design or a new implant material, finite element analysis (FEA) gives us a clear picture of what will be the patient bone response when an implant will be placed and loaded. FEA also gives us an opportunity to test a new implant material without putting a patient at risk. In this study, four different types of bone were incorporated with two different implant collar designs. Each implant assembly was subjected to vertical as well as oblique forces. The response of each bone type for the titanium alloy implant was recorded. A color-coded response for the magnitude and the location of the maximum stress received by the bone was observed. Maximum stresses were seen in the crestal region. As this is a computer-based model, dynamic loading was not possible. This study provided us with the possible outcome in patients under a static load. Further studies can be conducted in vivo to record dynamic loading responses as well as long-time loading responses.
Related Concept Videos
Stress Concentrations in Circular Shafts
Stress Concentrations
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Stress: General Loading Conditions
The shearing force, possessing potential directionality within the plane of the section, is simplified into two component forces running parallel to the x and y axes....
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each...
Collar Bearings