The "pipe-anchor" technique: A method for intra-aneurysmal loop reduction using a partially unsheathed pipeline stent
Omer Doron1, Justin Vranic1, Robert Regenhardt1
1Department of Neurosurgery, Massachusetts General Hospital, Harvard Medical School, Boston, MA, USA.
Summary
A novel "pipe anchor" technique uses a partially deployed pipeline stent to stabilize microcatheters for treating large aneurysms. This method simplifies endovascular navigation and stent deployment, offering a valuable approach for neurosurgeons.
Area of Science:
- Endovascular neurosurgery
- Cerebrovascular diseases
- Medical device technology
Background:
- Large aneurysms pose challenges for endovascular navigation, often requiring complex techniques like the "around-the-world" maneuver for distal access.
- Traditional methods for stabilizing microcatheters in such cases may involve additional devices and complex exchanges.
Purpose of the Study:
- To describe a novel technique utilizing a pipeline stent for microcatheter stabilization during endovascular treatment of large aneurysms.
- To evaluate the feasibility and efficacy of the "pipe anchor" technique in achieving stable microcatheter positioning and successful stent deployment.
Main Methods:
- The "pipe anchor" technique involves partial deployment of a pipeline stent distal to the aneurysm to create an anchoring point.
- The microcatheter is stabilized by the partially deployed stent, allowing for gradual unsheathing and straightening within the aneurysm.
- This technique facilitates loop reduction and alignment of the microsystem with the inflow and outflow vessels for stent deployment.
Main Results:
- The technique was successfully applied in two patients with large cavernous segment aneurysms.
- Pipeline stents were deployed using a Phenom 0.027" microcatheter without thromboembolic complications.
- Follow-up imaging confirmed good stent apposition and contrast stagnation, indicating successful treatment.
Conclusions:
- The "pipe anchor" technique effectively uses the radial force of a partially deployed pipeline stent for anchoring and microcatheter stabilization.
- This method simplifies the procedure by eliminating the need for additional devices or complex exchange maneuvers.
- The technique is a valuable addition to the endovascular neurosurgeon's armamentarium for treating select large aneurysms.