Noninvasive Assessment of Arterial Pulse-Pressure Variation During General Anesthesia: Clinical Evaluation of a New
Philippe Conter1, Josef Briegel1, Torsten Baehner2
1Department of Anesthesiology, University Hospital, LMU Munich, Munich, Germany.
Objectives:
To compare noninvasive pulse-pressure variation (PPV) measurements obtained from a new high-fidelity upper arm cuff using a hydraulic coupling technique to corresponding intraarterial PPV measurements.
Design:
The authors used prospective multicenter comparison and development studies for the new high-fidelity upper arm cuff.
Setting:
The study was performed in the departments of Anesthesiology at the Ludwig-Maximilians-Universität München Hospital, the University Hospital of Bonn, and the RoMed Hospital in Rosenheim (all Germany).
Participants:
A total of 153 patients were enrolled, undergoing major abdominal surgery or neurosurgery with mechanical ventilation. For the evaluation of PPV, 1,467 paired measurements in 107 patients were available after exclusion due to predefined quality criteria.
Interventions:
Simultaneous measurements of PPV were performed from a reference femoral arterial catheter (PPVref) and the high-fidelity upper arm cuff (PPVcuff). The new device uses a semirigid conical shell. It incorporates a hydraulic sensor pad with a pressure transducer, leading to a tissue pressure-pulse contour with all characteristics of an arterial- pulse contour.
Measurements And Main Results:
The comparative analysis of the included measurements showed that PPVref and PPVcuff were closely correlated (r = 0.92). The mean of the differences between PPVref and PPVcuff was 0.1 ± 2.0%, with 95% limits of agreement between -4.1% and 3.9%. To track absolute changes in PPV >2%, the concordance rate between the 2 methods was 93%.
Conclusions:
The new high-fidelity upper arm cuff method provided a clinically reliable estimate of PPV.
More Related Videos
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: