Technical failure rates for biometry between swept-source and older-generation optical coherence methods: a review
Piotr Kanclerz1,2, Idan Hecht3,4,5, Raimo Tuuminen3,6
1Department of Ophthalmology, Hygeia Clinic, ul. Jaśkowa Dolina 57, Gdańsk, 80-286, Poland. p.kanclerz@gumed.edu.pl.
Swept-source optical coherence tomography (SS-OCT) biometry shows a significantly lower technical failure rate (TFR) compared to partial coherence interferometry (PCI) and low-coherence optical reflectometry (LCOR) methods. This finding is crucial for improving accuracy in cataract and refractive surgery.
Area of Science:
- Ophthalmology
- Medical Optics
- Surgical Technology
Background:
- Precise ocular measurements are essential for successful cataract surgery and refractive lens exchange.
- Current biometry devices utilize various optical methods, including partial coherence interferometry (PCI), low-coherence optical reflectometry (LCOR), and swept-source optical coherence tomography (SS-OCT).
- SS-OCT employs longer wavelengths for enhanced penetration through dense ocular media, potentially improving measurement reliability.
Approach:
- A systematic literature search was conducted on PubMed and Scopus for studies comparing optical biometry methods.
- Included studies focused on patients undergoing routine cataract surgery and utilized at least two optical methods (PCI/LCOR vs. SS-OCT) within the same patient cohort.
- A meta-analysis was performed to pool and compare the technical failure rates (TFR) across different biometry technologies.
Key Points:
- Fourteen studies involving 2,459 eyes were analyzed, revealing an overall TFR of 5.47%.
- Significant differences in TFR were observed among the methods: PCI (15.72%), LCOR (6.88%), and SS-OCT (1.51%).
- The pooled TFR for infrared methods (PCI and LCOR) was 11.12%, substantially higher than SS-OCT's 1.51% (p < 0.001).
Conclusions:
- SS-OCT biometry demonstrates a significantly lower technical failure rate compared to PCI and LCOR methods.
- This reduction in TFR suggests SS-OCT offers improved reliability for ocular measurements in routine ophthalmic procedures.
- The findings support the adoption of SS-OCT technology for enhanced precision in pre-operative planning for cataract and refractive surgery.