Technical failure rates for biometry between swept-source and older-generation optical coherence methods: a review and meta-analysis
Related Experiment Video
Updated: Aug 1, 2025

Doppler Optical Coherence Tomography of Retinal Circulation
Published on: September 18, 2012
Piotr Kanclerz1,2, Idan Hecht3,4,5, Raimo Tuuminen3,6
1Department of Ophthalmology, Hygeia Clinic, ul. Jaśkowa Dolina 57, Gdańsk, 80-286, Poland. p.kanclerz@gumed.edu.pl.
Purpose:
Precise ocular measurements are fundamental for achieving excellent target refraction following both cataract surgery and refractive lens exchange. Biometry devices with swept-source optical coherence tomography (SS-OCT) employ longer wavelengths (1055-1300 nm) in order to have better penetration through opaque lenses than those with partial coherence interferometry (PCI) or low-coherence optical reflectometry (LCOR) methods. However, to date a pooled analysis showing the technical failure rate (TFR) between the methods has not been published. The aim of this study was to compare the TFR in SS-OCT and in PCI/LCOR biometry.
Methods:
PubMed and Scopus were used to search the medical literature as of Feb 1, 2022. The following keywords were used in various combinations: optical biometry, partial coherence interferometry, low-coherence optical reflectometry, swept-source optical coherence tomography. Only clinical studies referring to patients undergoing routine cataract surgery, and employing at least two (PCI or LCOR vs. SS-OCT) optical methods for optical biometry in the same cohort of patients were included.
Results:
Fourteen studies were included in the final analysis, which presented results of 2,459 eyes of at least 1,853 patients. The overall TFR of all included studies was 5.47% (95% confidence interval [CI]: 3.66-8.08%; overall I2 = 91.49%). The TFR was significantly different among the three methods (p < 0.001): 15.72% for PCI (95% CI: 10.73-22.46%; I2 = 99.62%), 6.88% for LCOR (95% CI: 3.26-13.92%; I2 = 86.44%), and 1.51% for SS-OCT (95% CI: 0.94-2.41%; I2 = 24.64%). The pooled TFR for infrared methods (PCI and LCOR) was 11.12% (95% CI: 8.45-14.52%; I2 = 78.28%), and was also significantly different to that of SS-OCT: 1.51% (95% CI: 0.94-2.41%; I2 = 24.64%; p < 0.001).
Conclusions:
A meta-analysis of the TFR of different biometry methods highlighted that SS-OCT biometry resulted in significantly decreased TFR compared to PCI/LCOR devices.