Vibrational characterization of cavitation in left ventricular assist device
Ditte Juhl Kisendal1,2, Katrine Rafn Josiasen1,2, Marie Nørkjaer Larsen1,2
1Department of Health Science and Technology, Aalborg University, Aalborg, Denmark.
Background:
The left ventricular assist device (LVAD) is a mechanical circulatory support device for patients with severe heart failure. Microbubbles caused by cavitation in the LVAD can potentially lead to physiological and pump-related complications. The aim of this study is to characterize the vibrational patterns in the LVAD during cavitation.
Methods:
The LVAD was integrated into an in vitro circuit and mounted with a high-frequency accelerometer. Accelerometry signals were acquired with different relative pump inlet pressures ranging from baseline (+20 mmHg) to -600 mmHg in order to induce cavitation. Microbubbles were monitored with dedicated sensors at the pump inlet and outlet to quantify the degree of cavitation. Acceleration signals were analyzed in the frequency domain to identify changes in the frequency patterns when cavitation occurred.
Results:
Significant cavitation occurred at the low inlet pressure (-600 mmHg) and was detected in the frequency range between 1800 and 9000 Hz. Minor degrees of cavitation at higher inlet pressures (-300 to -500 mmHg) were detected in the frequency range between 500-700, 1600-1700 Hz, and around 12 000 Hz. The signal power of the dominating frequency ranges was statistically significantly different from baseline signals.
Conclusion:
Vibrational measurements in the LVAD can be used to detect cavitation. A significant degree of cavitation could be detected in a wide frequency range, while minor cavitation activity could only be detected in more narrow frequency ranges. Continuous vibrational LVAD monitoring can potentially be used to detect cavitation and minimize the damaging effect associated with cavitation.
More Related Videos
07:41Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
03:40Point-of-Care Ultrasound for Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation Without Left Ventricular Venting
Published on: January 17, 2025
Related Concept Videos
Aortic Regurgitation I: Introduction
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mitral Regurgitation I: Introduction