Design of ear-contactless stethoscope and improvement in the performance of deep learning based on CNN to classify

Tanmay Sinha Roy1, Joyanta Kumar Roy2, Nirupama Mandal3

  • 1Department of Electrical Engineering, Haldia Institute of Technology, Haldia, West Bengal, India. tanmoysinha.roy@gmail.com.

Insights

A novel contactless stethoscope and deep learning models were developed for accurate heart sound classification. The CNN-based Inception Net achieved 99.65% accuracy, outperforming other models for diagnosing valvular heart conditions.

Area of Science:

  • Biomedical Engineering
  • Artificial Intelligence in Healthcare
  • Cardiology

Background:

  • Cardiac-related disorders are a growing global health concern, necessitating accurate diagnostic tools.
  • The COVID-19 pandemic highlighted challenges in traditional auscultation due to infection control measures, driving the need for contactless methods.
  • Accurate classification of cardiovascular diseases is crucial for timely and effective patient management.

Purpose of the Study:

  • To design a low-cost, contactless stethoscope for remote auscultation of heart sounds.
  • To enhance the performance of deep learning models, specifically Recurrent Neural Networks (RNN) and Convolutional Neural Networks (CNN), for classifying valvular heart diseases.
  • To optimize deep learning model performance through hyperparameter tuning for real-time analysis.

Main Methods:

  • Development of a contactless stethoscope utilizing a Bluetooth-enabled micro-speaker.
  • Implementation and comparison of deep learning models (CNN, RNN, hybrid CNN-RNN, LSTM) using acoustic, time, and frequency domain features.
  • Hyperparameter tuning (learning rate, dropout rate, hidden layers) to optimize model performance and learning curves.
  • Validation using heart sound recordings from normal and diseased patients from a standard data repository.

Main Results:

  • The proposed CNN-based Inception Network model achieved a high accuracy of 99.65% ± 0.06%, with 98.8% ± 0.05% sensitivity and 98.2% ± 0.19% specificity.
  • A hybrid CNN-RNN architecture attained 91.17% ± 0.03% accuracy after hyperparameter optimization.
  • The LSTM-based RNN model achieved 82.32% ± 0.11% accuracy.
  • The CNN-based Inception Net demonstrated superior performance compared to other machine learning algorithms and deep learning models evaluated.

Conclusions:

  • The developed contactless stethoscope offers a viable alternative for heart sound auscultation, particularly in situations requiring infection control.
  • Optimized deep learning models, especially the CNN-based Inception Network, show significant potential for accurate and automated classification of valvular heart diseases.
  • This approach facilitates improved real-time analysis and diagnosis of cardiovascular conditions, contributing to advancements in remote healthcare monitoring.