Evaluating the Clinical Accuracy of a Non-invasive Single-Fasting-Calibration Glucometer in Patients with Diabetes: A
Ang Li1, Xiang Li2, Yuanmeng Xu2
1Department of Endocrinology, Peking University First Hospital, Beijing, 100034, China.
Introduction:
The aim of this study was to evaluate the stability and accuracy of glucose measurements determined using the metabolic heat conformation (MHC)-based non-invasive glucometer in a multicentre, self-controlled clinical trial. This device is the first to obtain a medical device registration certificate awarded by the National Medical Products Administration of China (NMPA).
Methods:
The multicentre clinical study was conducted at three sites and enrolled 200 subjects whose glucose was measured with a non-invasive glucometer (the Contour Plus blood glucose monitoring system) and by venous plasma glucose (VPG) measurements, in a fasted state and at 2 and 4 h after meals.
Results:
Based on both the non-invasive and VPG measurements, 93.9% (95% confidence interval 91.7-95.6%) of the blood glucose (BG) values fell within consensus error grid (CEG) zones A + B. The measurements obtained in a fasted state and at 2 h after meals were more accurate, with 99.0% and 97.0% of the BG values, respectively, falling within zones A + B. Compared to those subjects who received insulin, the proportion of values in zones A + B and the correlation coefficients were 3.1% and 0.0596 higher, respectively. The accuracy of the non-invasive glucometer was influenced by the level of insulin resistance calculated by the homeostatic model assessment method, which had a correlation coefficient with the mean absolute relative difference of - 0.1588 (P = 0.0001).
Conclusion:
The MHC-based non-invasive glucometer assessed in the present study demonstrates generally high stability and accuracy in the glucose monitoring of people with diabetes. The calculation model needs to be further explored and optimised for patients with different diabetes subtypes, levels of insulin resistance and insulin secretion capacity.
Clinical Trial Registry Number:
ChiCTR1900020523.
More Related Videos
08:30Intraperitoneal Glucose Tolerance Test, Measurement of Lung Function, and Fixation of the Lung to Study the Impact of Obesity and Impaired Metabolism on Pulmonary Outcomes
Published on: March 15, 2018
08:13Study of In Vivo Glucose Metabolism in High-fat Diet-fed Mice Using Oral Glucose Tolerance Test OGTT and Insulin Tolerance Test ITT
Published on: January 7, 2018
Related Concept Videos
Hypoglycemia and Glucagon
Diabetes: Symptoms, Diagnosis, and Complications
Diabetes Mellitus: Type 2 and Gestational
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...