Related Experiment Video
Updated: Aug 1, 2025

Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Safety and Effectiveness of Expandable Intravertebral Implant Use for Thoracolumbar Burst Fractures
Haokang Wei1, Daniel Hsu1, Himanshu Katta2
1Department of Radiology, Rutgers Robert Wood Johnson Medical School, New Brunswick, New Jersey.
Purpose:
To determine the safety and effectiveness of an expandable intravertebral implant (Spinejack; Stryker, Kalamazoo, Michigan) as a treatment option for patients with thoracolumbar spine burst fractures without fracture-related neurologic deficit.
Materials And Methods:
Imaging studies before and after expandable intravertebral implantation and medical records of 33 patients, 11 (33.3%) men and 22 (66.6%) women with an overall mean age of 71.7 years ± 8.3, were reviewed for 60 thoracolumbar Magerl Type A3 injuries secondary to osteoporosis, trauma, or malignancy. The mean follow-up time was 299 days.
Results:
Implantation of an expandable intravertebral device resulted in a statistically significant reduction in bone fragment retropulsion (mean ± SD, 0.64 mm ± 16.4; P < .001), reduction in the extent of canal compromise (mean, 5.5%; P < .001), increased central canal diameter (mean ± SD, 0.71 mm ± 1.3; P < .001), and restoration of vertebral body height, with a mean increase of 5.0 mm (P < .001). However, the implantation did not result in a statistically significant kyphosis reduction (mean, 1.38°; P = .10). All patients except for 1 reported improvement in pain after surgery, with a mean improvement of 1.54 on a 4-point pain scale (P < .001). No clinically significant adverse events were reported.
Conclusions:
This study suggests that expandable intravertebral device implantation is a safe and effective treatment for thoracolumbar vertebral burst fractures in patients without fracture-related neurologic deficit. Although implantation did not result in a statistically significant reduction in kyphotic angle, it offered significant improvement in pain, vertebral body height, fracture fragment retropulsion, and central canal diameter compromise.