Related Experiment Video
Updated: Aug 1, 2025

In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
Survival of Patient-Specific Unicondylar Knee Replacement
Patrick Weber1,2,3, Melina Beck1, Michael Klug4,5
1ECOM, Arabellastraße 17, 81925 München, Germany.
Abstract:
Unicompartmental knee arthroplasty (UKA) in isolated medial or lateral osteoarthritis leads to good clinical results. However, revision rates are higher in comparison to total knee arthroplasty (TKA). One reason is suboptimal fitting of conventional off-the-shelf prostheses, and major overhang of the tibial component over the bone has been reported in up to 20% of cases. In this retrospective study, a total of 537 patient-specific UKAs (507 medial prostheses and 30 lateral prostheses) that had been implanted in 3 centers over a period of 10 years were analyzed for survival, with a minimal follow-up of 1 year (range 12 to 129 months). Furthermore, fitting of the UKAs was analyzed on postoperative X-rays, and tibial overhang was quantified. A total of 512 prostheses were available for follow-up (95.3%). Overall survival rate (medial and lateral) of the prostheses after 5 years was 96%. The 30 lateral UKAs showed a survival rate of 100% at 5 years. The tibial overhang of the prosthesis was smaller than 1 mm in 99% of cases. In comparison to the reported results in the literature, our data suggest that the patient-specific implant design used in this study is associated with an excellent midterm survival rate, particularly in the lateral knee compartment, and confirms excellent fitting.