Related Experiment Video
Updated: Aug 1, 2025

Author Spotlight: Rapid Prototyping and Testing of Self-Expanding Nitinol Frames for Transcatheter Implantable Devices
Published on: March 7, 2025
Relevant Choices Affecting the Fatigue Analysis of Ni-Ti Endovascular Devices
Francesca Berti1, Alma Brambilla2, Giancarlo Pennati1
1Department of Chemistry, Materials and Chemical Engineering "G. Natta" (LaBS), Politecnico di Milano, 20133 Milan, Italy.
Abstract:
Ni-Ti alloys are widely used for biomedical applications due to their superelastic properties, which are especially convenient for endovascular devices that require minimally invasive insertion and durable effects, such as peripheral/carotid stents and valve frames. After crimping and deployment, stents undergo millions of cyclic loads imposed by heart/neck/leg movements, causing fatigue failure and device fracture that can lead to possibly severe consequences for the patient. Standard regulations require experimental testing for the preclinical assessment of such devices, which can be coupled with numerical modeling to reduce the time and costs of such campaigns and to obtain more information regarding the local state of stress and strain in the device. In this frame, this review aimed to enlighten the relevant choices that can affect the outcome of the fatigue analysis of Ni-Ti devices, both from experimental and numerical perspectives.
Related Concept Videos
Fatigue
Node Analysis for AC Circuits
To unravel the complexities of this system, nodal analysis is employed, a powerful technique founded on Kirchhoff's current law (KCL), which remains valid for phasors. AC circuits can effectively be...