Rereferencing of clinical EEGs with nonunipolar mastoid reference to infinity reference by REST
Li Dong1, Yongxiu Lai2, Mingjun Duan3
1The Clinical Hospital of Chengdu Brain Science Institute, MOE Key Lab for Neuroinformation, School of Life Science and Technology, University of Electronic Science and Technology of China, Chengdu, China; Research Unit of NeuroInformation, Chinese Academy of Medical Sciences, 2019RU035 Chengdu, China; Sichuan Institute for Brain Science and Brain-Inspired Intelligence, Chengdu 611731, China.
Objective:
Conventional electroencephalography (EEG) offline subtraction rereferencing is invalid for many clinical practices when adopting a specific nonunipolar recording montage (e.g., the ipsilateral mastoid (IM) and contralateral mastoid (CM)). Further comparative analyses would thus be blocked due to the lack of a uniform offline reference. Therefore, our goal was to resolve this problem by introducing and assessing the reference electrode standardization technique (REST) to transform nonunipolar mastoid montages into a computational zero reference at infinity (IR) offline.
Methods:
For EEG signals and power/connectivity configurations, simulation and clinical schizophrenia resting-state EEG datasets were used to investigate the performance of REST.
Results:
REST produced small absolute errors (signal level: 1.21-1.26; power: 0.0057-0.021; connectivity: 0.066-0.088) and high correlations (>0.9) between the IM/CM-IR and true IR references. Using clinical data with the IM online reference, REST revealed valuable changes in spectral and connectivity (P < 0.05) in schizophrenia patients, consistent with previous studies.
Conclusions:
These results demonstrated that REST transformation could be adopted to resolve the offline rereferencing of clinical EEGs with specific nonunipolar mastoid references.
Significance:
REST could be an effective and robust resolution for nonunipolar clinical EEGs and could therefore retrieve these data for further analysis by deriving a favorable offline reference IR.